A combined optical and electrical textile for non-invasive physiological monitoring

From Applied Optics Wiki
Revision as of 16:52, 27 May 2016 by Chong LIU (talk | contribs) (Created page with "'''Introduction''' •Peripheral capillary oxygen saturation (SpO2), a very accurate and widely used estimation of blood oxygen saturation, can be calculated from two different w...")

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Introduction •Peripheral capillary oxygen saturation (SpO2), a very accurate and widely used estimation of blood oxygen saturation, can be calculated from two different wavelength PPG signals.
•Photoplethysmogram(PPG) is a term referring the changes in the blood volume caused by heart pumps (with each cardiac cycle).
•The blood volume changes can be detected by using a LED to illuminate light to the skin and then measuring the reflected or transmitted light intensity.
•Utilising optical fibres incorporated into a textile to measure the SpO2, the pulse oximeter is available to wear.



Working Principle •There are two different types of haemoglobin (oxyhaemoglobin(HbO2) and deoxygenated haemoglobin(Hb)). SpO2 is the fraction of HbO2 to total haemoglobin(HbO2+Hb)


•Since two different types of haemoglobin have different optical absorption performance(shown in Fig.1), SpO2 can be deduced from the PPG recorded at two different wavelengths.


•The device is developed to collect reflected PPG signals to measure the SpO2(shown in Fig.2).


•Optical fibres are applied to illuminate two different wavelengths light to skin and collect reflected light.


•Two different wavelength light used in the device are 625nm(red) and 850nm(NI)



$7WD 0R@C%YQKY7TSK%SUQB.pngFile:Example.jpg

Experiment Results

File:Example.jpg

File:Example.jpg File:Example.jpg